GD***1B Semi-auto PMCC Flash Point Tester
Purpose
The instrument is designed and made as per the National
Standard GB/T**1 “Test Methods for Flash Point of Petroleum
Products (Closed Cup Methods)”. It is used to determine the lowest
temperature of a petroleum product sample when the mixture of
sample vapor and air around the sample contacts and flashes when
heated in a closed cup under stipulated conditions. The lowest
temperature is the flash point of the sample.
Adopt microprocessor techniques and equipped with advanced
software, it can control temperature, stir sample, ignite, detect,
and print out test data automatically. It has a six digits
display.
Main parameters
(I) Operation condition
1. Ambient temperature: **0℃~**0℃;
2. Relative humidity: < *0%;
3. Power supply: AC **0V±*0%, *0±1Hz;
(II) Temperature measurement
1. Measurement range: ambient temperature~**0℃;
2. Accuracy: 0.2%;
3. Resolution: 0.1 ℃;
4. Repeatability:
(1) Flash Point is lower than **4 ℃: error is 2 ℃;
(2) Flash Point is higher than **4 ℃: error is 4 ℃;
(III) Basic parameters
1. Interval for flame application
(1) Flash point is lower than **4 ℃: 1 ℃;
(2) Flash point is higher than **4 ℃: 2 ℃;
2. Temperature rising rate: as per GB/T*****3;
3. Stirring speed: **5 RPM
4. Ignition mode: by gas flame (4 mm);
5. Total power consumption: not more than **0 W;
6. Dimension:
(1) Main case: **0×**0×**0 (mm);
(2) Measurement case: **0×**0×**0 (mm);
7. Net weight: *0 kg;