Abstract
1.1 Applications
Type YX***0 X-ray crystal orientation unit is designed on the principle of x-ray diffraction. It is a indispensable instrument for precise orientating of various natural and artificial single crystals (such as piezo-crystal optical crystal laser crystal, and semiconductor crystal ECT)
it Is can used in above-mentioned of different crystal orientation cut , it is make crystal device indispensable instrument to process set with cutting machine.
1.2 Features
A).Easy for Operation.
B).Digital display for angles.
C).Display may be reset at any angle, for user setting crystal angle.
D).Special integrater is used for peak enlargement to increase accuracy.
E).Modularized circuit makes maintenance easier.
F).Reasonable structure design, to protect the user from x-ray injury.
G).Sample with different size can be measured with various attachments.
(It depends on different needs)
2. Technical parameter
2.1 Environment condition
(1) Elevation is lower than ***0m
(2) Environment temperature :*2°*-**0°C
(3) Relativity humidity≤*0%
(4) Power: AC **0V, 5A, voltage undulations: ≤±*0%, power capacity:≥0.5KVA
(5) Good grounding device, grounding resistance≤4Ω
(6) Power supply line must have no electric welding machine and so on, which can make interference in high frequency and electric arc.
2.2 Specification
(1) X-ray generator:
X-ray tube:
Cu target, anode-grounding, fan cooling;
Tube voltage: *0KVP
Tube current: 0~5mA, adjusted successively;
(2) Total power: ≤0.3KW
(3) Sensitivity: excel to *0" (standard sio2 ***1)
(4) Goniometer :
a). The total error≤*0
b) Angle Reading 3
c).Counter angle:**0°*-***0°
d).Sample angle: **0°*-**0°
e). Digital reading: digital display of angle in deg.min.sec.and the accuracy 3"
(5 )Recording unit
a).Counter: JBIX Geiger counter and detector voltage **0~***0V
b).Time constant: 1, 2,3 steps
(6) External dimensions: ***0(W) ×**0(L) ×***0(H)
(7) Weight: **0kg
3.The principle
High voltage generator supply X-ray tube with high voltage, product x-ray, shoot on the sample, according to nλ=2dsinθ,product diffraction, and the digital counter absorb the x-ray and display through the uA meter.
4. Component and Features
4.1.YX***0 x-ray orientation unit is consisted of mechanical part, which made up with the table, tube housing , goniometer and various accessories; and electrical part ,which made up with voltage stabilizer and controller , high voltage transformer, magnifying, digital counter, air pump , cooling fan , x-ray tube.
4.2. The tube housing of orientation unit is horizontal and the cover is on the top , Easy to replace the x-ray tube.
4.3. Counter tube and samples share one scale, in the face of operator, easy to read.
4.4. Counter tube relative to the sample distance is adjustable at his arm, in order to meet and measure the need of the sample of different size.
4.5. Air pump have long performance life , it is convenient to maintain.
4.6. The switch of power, voltage of x-tube and the current knob are in the front of the tube housing.
4.7 Safety defence cover to avoid x-rays injury to the user.
5. Installation and adjusting
5.1. This unit is also need the following conditions:
(1) Place where be well-ventilated, and less dust.
(2) Place where flat and no shake.
5.2. Installation
This unit should be place in the room after open the packing box, the first item, you should check the parts according the packing list. if there are some problem, you should contact with the factory immediately .
(1) The preparation before installation
Should have the necessary tools for moving and install this unit.
Should clean the unit after open the packing box . clear the x-ray tube and the voltage cable .the body of the unit should be firmed.
(2) The installation of the x-ray tube and the voltage cable
Keep the x-ray tube and the voltage cable clean ,first loose the screws over the cover , then move the cover ,and loose the screws on the nipM3,and make the x-ray tube and high voltage cable become a whole. put the x-ray tube in tube housing .
Caution: Dont contact with the Be window of X-ray tube during the installation .dont link the high voltage cable temporarily with high voltage transformer.
(1) Counter installation
First check it carefully, install this unit on the arm at goniometer, the distance is the shorten the best between this unit and center of the sample , connect the cable plug with the outlet.
(2) Cable connect
a).The first you must discharge the connector on the high voltage transformer with earth!!
and then connect the high voltage cable with the high voltage transformer, with screw (M5)
b) Connect the power cable with D1, D2 (big hole)
c).This unit must has earth connection . (smaller hole)
5.3. Testing.
On the basis of the good installation . Switch on the main power , the indicator lamp will be light (red).the following should work well.
(1) X-ray tubes filament is light, adjust the mA knob, at the same time, the brightness of filament should change.
(2) The cooling fan should work and make voice.
(3) Switch on the air pump, the sample plan can attract the standard sample.
(4) Push ON button of the high-voltage, the high-voltage indicator lamp is light (red), Open the shutter of X-ray , there is X-ray penetrates. In order to check whether there is X ray that penetrates, you can adopt the fluorescence board or through the partly display of the counter and uA meter
Under the condition of the X-ray penetrates and normal work of partly display,
Calibration goniometer as follows,
a) Put standard quartz (***1 ) at the sample plan,
b) Unclamp the fastening hand knob of the detector, move the detector to reach *6°*0'(Examine the angle scale indicator on the angle appearance through observing), then lock the hand knob.
c) Rotate the hand wheel and transfer the sample to *3°*0, regulate every knob of the register in the best position, Switch on the high-voltage and open X-ray shutter, slightly rotate the hand wheel until intensity is highest on uA meter indicator, at this time you can preset the digital switch to the *3.*0.0,and then press the reset key, or you can set the digital switch to the *0.*0.0, if you like then press the reset key.